Using Computational Fluid Dynamics to Examine the Hemodynamics of Ar tificial Heart Valves

Login or register to view PDF.
Citation
US Cardiology 2004;2004:1(1):1-5

Pages

Background

Since the first successful implantation of a prosthetic heart valve four decades ago, over 50 different designs have been developed including both mechanical and bioprosthetic valves. Today, with over 200,000 implants worldwide each year, the most widely implanted design is the mechanical bileaflet prosthesis. Several other mechanical valves are currently available and many of them have good bulk forward flow hemodynamics. However, mechanical valve implants suffer from complications resulting from thrombus deposition and patients implanted with these valves need to be under long-term anti-coagulant therapy. In general blood thinners are not needed with bioprosthetic implants, but tissue valves suffer from structural failure with an average life-time of 10-12 years before replacement is needed.

Flow-induced stresses on the formed elements in blood have been implicated in thrombus initiation within the mechanical valve prostheses. Regions of stress concentration on the leaflets during the complex motion of the leaflets have been implicated with structural failure of the leaflets with bioprosthetic valves. In vivo and in vitro experimental studies have yielded valuable information on the relationship between hemodynamic stresses and the problems associated with the implants. More recently, Computational Fluid Dynamics (CFD) has emerged as a promising tool, which, alongside experimentation, can yield insights of unprecedented detail into the hemodynamics of prosthetic heart valves. For CFD to realize its full potential, however, it must rely on numerical techniques that can handle the enormous geometrical complexities of prosthetic devices with spatial and temporal resolution sufficiently high to accurately capture all hemodynamically relevant scales of motion. Such algorithms do not exist today and their development is at the forefront of ongoing research. For CFD to further gain the confidence of valve designers and medical practitioners it must also undergo comprehensive validation with experimental data. Such validation requires the use of high-resolution flow measuring tools and techniques and the integration of experimental studies with CFD modeling.

The link between hemodynamics and clinical complications

Platelets can become hyper-activated due to shear forces and they present a risk for the development of a thrombotic event. As early as 1970, Harker and Slichter13 showed that patients with first-generation mechanical valves such as the ball-and-cage and tilting-disc had a shortened platelet half-life due to increased incidence of platelet destruction and activation.

Pages

References
  1. Baldwin J,Tarbell J, Deutsch S et al.,"Mean Velocities and Reynolds Stresses Within Regurgitant Jets Produced by Tilting Valves",American Society of Artifical Internal Organs Transactions, (1991), 37
    PubMed
  2. Chandran K B, Cabell G N, Khalighi B et al., "Laser anemometry measurements of pulsatile flow past aortic valve prostheses", J. Biomech. (1983), 16 (10).
    Crossref | PubMed
  3. Chandran K B, Khalighi B and Chen C J, "Experimental study of physiological pulsatile flow past valve prosthesis in a model of human aortaÔÇöII.Tilting disc valves and the effect of orientation", J. Biomech. (1985), 18 (10): pp. 773-780.
    Crossref | PubMed
  4. Ellis J T,"An In Vitro Investigation of the Leakage and Hinge flow fields through bileaflet mechanical heart valves and their relevance to thrombogenesis, PhD thesis", Georgia Institute of Technology, (1999).
  5. Ellis J T, Healy T M, Fontaine A A et al.,"Velocity measurements and flow patterns within the hinge region of a Medtronic Parallel bileaflet mechanical valve with clear housing", J. Heart Valve Dis. (1996), 5 (6).
    PubMed
  6. Ellis J T, Healy T M, Fontaine A A et al.,"An in vitro investigation of the retrograde flow fields of two bileaflet mechanical heart valves", J. Heart Valve Dis. (1996); 5 (6).
    PubMed
  7. Ellis J T,Travis B R and Yoganathan A P. "An in vitro study of the hinge and near-field forward flow dynamics of the St. Jude Medical Regent bileaflet mechanical heart valve", Ann. Biomed. Eng. (2000), 28 (5).
    Crossref | PubMed
  8. Ellis J T and Yoganathan A P,"A comparison of the hinge and near-hinge flow fields of the St Jude medical hemodynamic plus and regent bileaflet mechanical heart valves", J.Thorac. Cardiovasc. Surg. (2000), 119 (1).
    Crossref | PubMed
  9. Fareed J, Bick R, Squillaci G et al., "Molecular Markers of Hemostatic Disorders: Implications in the Diagnosis and Therapeutic Management of Thrombotic and Bleeding Disorders", Clinical Chemistry (1983), 29.
    PubMed
  10. Fontaine A, He S, Stadter R et al., "In Vitro Assessment of Prosthetic Valve Function in Mitral Valve Replacement with Chordal Preservation Techniques", Journal of Heart Valve Disease (1996), 5.
    PubMed
  11. Ge L., Jones S C, Sotiropoulos F et al.,"Numerical Simulation of Flow in Mechanical Heart Valves: Grid Resolution and Flow Symmetry", ASME J. of Biomechanical Eng. (2003), 125 (5): pp. 709-718.
    Crossref | PubMed
  12. Gross J M, Shu M C S, Dai F F et al., "A Microstructural Flow Analysis within a Bileaflet Mechanical Heart Valve Hinge", Journal of Heart Valve Disease (1996), 5 (6).
    PubMed
  13. Harker L and Slichter S,"Studies of Platelet and Fibrinogen Kinetics in Patients with Prosthetic Heart Valves", The New England Journal of Medicine (1970), 283.
    Crossref | PubMed
  14. Hsu A T,Yun J X and Hwang N H C, "Application of an Unstructured Grid Algorithm to Artificial Heart Valve Simulations", ASAIO Journal (1999), 45.
    Crossref | PubMed
  15. King M J, Corden J, David T et al., "A Three-Dimensional, Time-Dependent Analysis of Flow Through a Bileaflet Mechanical Heart Valve: Comparison of Experimental and Numerical Results", Journal of Biomechanics (1996); 29.
    Crossref | PubMed
  16. King M J, David T and Fisher J, "An Initial Parametric Study on Fluid Flow Through Bileaflet Mechanical Heart Valves Using Computational Fluid Dynamics", Proceedings of the Institution of Mechanical Engineers-Part H (1994), 208.
    Crossref
  17. Kiris C, Kwak D, Rogers S et al., "Computational Approach for Probing the Flow Through Artificial Heart Devices", Journal of Biomechanical Engineering (1997), 119.
    Crossref | PubMed
  18. Koppensteiner R, Moritz A, Schlick W et al., "Blood Rheology After Cardiac Valve Replacement with Mechanical Prostheses or Bioprostheses", The American Journal of Cardiology (1991), 67.
    Crossref | PubMed
  19. Leo H L, He Z, Ellis J T et al., "Microflow fields in the hinge region of the CarboMedics bileaflet mechanical heart valve design", J.Thorac. Cardiovasc. Surg. (2002), 124 (3).
    Crossref | PubMed
  20. Lip G, Lowe G, Rumley A et al.,"Increased Thrombogenesis Markers of in Chronic Atrial Fibrillation: Effects of Warfarin Treatment", British Heart Journal (1995), 73.
    Crossref | PubMed
  21. Makhijani V B, Siegel J M and Hwang N H C, "Numerical Study of Squeeze-Flow in Tilting Disc Mechanical Heart Valves", Journal of Heart Valve Disease (1996), 5.
    PubMed
  22. Meyer R S, Deutsch S, Maymir J C et al., "Three-component laser Doppler velocimetry measurements in the regurgitant flow region of a Bjork-Shiley monostrut mitral valve", Ann. Biomed. Eng. (1997), 25 (6).
    Crossref | PubMed
  23. Myhre E, Hellem A, Stormorken H et al., "Intravascular Hemolysis, Platelet Consumption and Platelet Adhesiveness in Patients with Prosthetic Heart Valves", Scandinavian Journal of Thoracic Cardiovascular Surgery (1971),5.
    Crossref | PubMed
  24. Shim E and Chang K, "Three-Dimensional Vortex Flow Past a Tilting Disc Valve Using a Segregated Finite Element Scheme", Computational Fluid Dynamics Journal (1994), 3.
  25. Walenga J and Pifarre R, "Hemostatic Alterations in Myocardial Infarction and Use of Newer Laboratory Methods in Diagnosis and Prognosis", Cardiac Surgery: State of the Art Reviews (1992), 6.
  26. Walker P and Yoganathan A P,"In Vitro Pulsatile Flow Hemodynamics of Five Mechanical Aortic Heart Valve Prostheses", European Journal of Cardiothoracic Surgery (1992), 6 (Supp 1).
    Crossref | PubMed
  27. Yoganathan A P,Woo Y and Sung H, "Turbulent Shear Stress measurements in the Vicinity of Aortic Heart Valve Prostheses", Journal of Biomechanics (1986), 19.
    Crossref | PubMed